S1200 ICU Ventilator Machine

The ventilator is suitable for various kinds of medical institutions for cardiopulmonary resuscitation respiratory support. Acute respiratory failure caused by various reasons or incomplete oxygenation dysfunction. Indra-operation, post-operation respiratory support, other respiratory treatment.

Description

S1200 ICU Ventilator

ADULT · PEDIATRIC

Friendly Powerful Reliable

Application

The ventilator is suitable for various kinds of medical institutions for cardiopulmonary resuscitation respiratory support. Acute respiratory failure caused by various reasons or incomplete oxygenation dysfunction. Indra-operation, post-operation respiratory support, other respiratory treatment.

Features

  • l 15” TFT LCD touch screen displays the ventilation parameters, alarm information and waveforms.
  • Air compressor runs stably with low noise, creates quiet environment for doctors and patients.
  • Integration breathing circuit design.
  • Multiple working modes such as volume control and pressure limit.
  • Multiple parameters monitoring interface to show clear parameter to the operator, allow operator know the patient conditions in all aspects.
  • Real time pressure-time, flow-time loop graphics, pressure volume loop graphics and high precision ETCO2, O2 concentration detection function included.
  • Electrical PEEP control
  • Breathing valve is disassembly and can be disinfected.

Safety

  • Three levels alarm system, visual and sound alarm information.
  • With multiple type of alarms, reminding and protection functions.
  • Self-check before running, eliminate system malfunction.
  • Advanced power management control technology.
  • With built-in backup power supply, when normal power supply shut down, back-up power supply will start work automatically.
  • Separate design of electric and gas, keep safety running of ventilator.

Specifications

Ventilation modes: IPPV, A/C, PCV, SIMV, SPONT/CPAP, PSV
Ventilator parameter ranges
Tidal volume(Vt) 0 ~ 2000mL
Frequency (Freq)

Oxygen concentration

1bpm ~ 100 bpm

21%-100%

I/E 4:1~1:8
PEEP 0cmH2O ~ 40 cmH2O
Pressure Limit 20 cmH2O ~ 100 cmH2O
Monitored Parameters
Frequency (Freq) 0 /min ~ 100 /min
Tidal volume(Vt) 0 mL ~ 2500 mL
MV

Airway pressure

Dynamic lung compliance testing

ETCO2 Concentration

0 L/min ~ 99 L/min

0 cmH2O ~ 100 cmH2O

1mL/cmH2O ~ 1000mL/cmH2O

0mmHg ~ 152mmHg (0%-20%)

Oxygen concentration 15 % ~ 100 %
Oscillographs display:
P-T(pressure – time)
F-T(flow – time)
P-V loop (pressure – volume loop)
V-T(volume – time )
ETCO2-T (End-tidal CO2 – time)
Alarms and protection
The AC power failure alarm Power failure or no connection
Internal battery backup low voltage alarm ≤11.3±0.3V
No tidal volume No tidal volume within 6s
High Minute Volume alarm

Low Minute Volume alarm

5L/min-99L/min

1L/min-30L/min

High Airway pressure alarm

Low Airway pressure alarm

High oxygen concentration alarm

Low oxygen concentration alarm

20cmH2O-100cmH2O

0cmH2O-20cmH2O

19%-100%

18%-99%

Continuous pressure alarm (PEEP+1.5cmH2O) over 16s
Suffocation warning

Fan error

Oxygen deficit

                5-60s

Show on screen

Show on screen

The maximum limited pressure <12.5 kPa
Working conditions
Gas source O2, AIR
Pressure 280kPa-600kPa
Voltage ~220V±22V
Power frequency 50Hz ± 1Hz
Input power 900VA (with air compressor)

250VA (without air compressor)

Ventilation modes: IPPV, A/C, PCV, SIMV, SPONT/CPAP, PSV
Ventilator parameter ranges
Tidal volume(Vt) 0 ~ 2000mL
Frequence (Freq)

Oxygen concentration

1bpm ~ 100 bpm

21%-100%

I/E 4:1~1:8
PEEP 0cmH2O ~ 40 cmH2O
Pressure Limit 20 cmH2O ~ 100 cmH2O
Monitored Parameters
Frequence (Freq) 0 /min ~ 100 /min
Tidal volume(Vt) 0 mL ~ 2500 mL
MV

Airway pressure

Dynamic lung compliance testing

ETCO2 Concentration

0 L/min ~ 99 L/min

0 cmH2O ~ 100 cmH2O

1mL/cmH2O ~ 1000mL/cmH2O

0mmHg ~ 152mmHg (0%-20%)

Oxygen concentration 15 % ~ 100 %
Graphical display:
P-T(pressure – time)
F-T(flow – time)
P-V loop (pressure – volume loop)
V-T(volume – time )
ETCO2-T (End-tidal CO2 – time)
Alarm and protection
The AC power failure alarm Power failure or no connection
Internal battery backup low voltage alarm ≤11.3±0.3V
No tidal volume No tidal volume within 6s
High Minute Volume alarm

Low Minute Volume alarm

5L/min-99L/min

1L/min-30L/min

High Airway pressure alarm

Low Airway pressure alarm

High oxygen concentration alarm

Low oxygen concentration alarm

20cmH2O-100cmH2O

0cmH2O-20cmH2O

19%-100%

18%-99%

Continuous pressure alarm (PEEP+1.5cmH2O) over 16s
Suffocation warning

Fan error

Oxygen deficit

                5-60s

Show on screen

Show on screen

The maximum limited pressure <12.5 kPa
Working conditions
Gas source O2,AIR
Pressure 280kPa-600kPa
Voltage ~220V ±22V
Power frequency 50Hz ± 1Hz
Input power 900VA (with air compressor)

250VA (without air compreesor)